Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8600A: CAL ADJ /5100,5205 DATE: 08-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 105 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A STANDARD: Precision 1 MOhm Resistor ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Precision 1 MOhm Resistor 1.002 ASK- W 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Before performing adjustments allow the UUT to stabilize 1.005 DISP (5 min.) at 23+/-5degC. Humidity less than 80%. 1.006 DISP Remove the cover from the UUT. 1.006 DISP Turn the UUT ON and allow it to warm up for 15 minutes. 1.007 HEAD DC VOLTAGE CALIBRATION ADJUSTMENT 1.008 RNG 200mV 1.009 DISP Connect the precision 1 MOhm resistor across the UUT 1.009 DISP Volt-Ohm HI and Volt-Ohm LO terminals. 1.010 DISP Adjust C24 for a UUT reading of 00.00+/-0.01mV 1.011 DISP Remove the resistor. 1.012 DISP Connect the 5100B and UUT as follows: 1.012 DISP [32] 5100B OUTPUT HI to UUT V-Ohm HI 1.012 DISP [32] 5100B OUTPUT LO to UUT V-Ohm LO 1.013 5100 190.00mV S 2W 1.014 MEMI Enter UUT reading in millivolts: 1.015 MATH M[1] = MEM 1.016 5100 -190.00mV S 2W 1.017 MEMI Enter UUT reading in millivolts: 1.018 MATH MEM1 = M[1] + MEM 1.019 JMPZ 1.023 0.01U 1.020 MATH MEM = M[1] 1.021 DISP Adjust C30 for a UUT reading of [MEM]+/-0.01mV. 1.022 JMP 1.013 1.023 5100 * S 1.024 RNG 2V 1.025 5100 1.9000V S 2W 1.026 DISP Adjust R57 for a reading of 1.9000V or 1.9001V. 1.027 5100 -1.9000V S 2W 1.028 DISP Adjust R57 for a reading of -1.9000V or -1.9001V. 1.029 5100 * S 1.030 RNG 20V 1.031 5100 19.000V S 2W 1.032 DISP Adjust R4 for a reading of 19.000+/-0.001V. 1.033 5100 -19.000V S 2W 1.034 DISP Adjust R4 for a reading of -19.000+/-0.001V. 1.035 5100 * S 1.036 RNG 200V 1.037 5100 190.00V S 2W 1.038 DISP Adjust R6 for a reading of 190.00+/-0.01V. 1.039 5100 * S 1.040 RNG 1200V 1.041 5100 1000.0V S 2W 1.042 DISP Adjust R8 for a reading of 1000.0+/-0.1V. 1.043 5100 * S 1.044 HEAD OHMS CALIBRATION ADJUSTMENT 1.045 RNG 2000kZ 1.046 RESF 1000.0kZ S 2W 1.047 DISP Adjust R21 for a reading of 1000.0+/-0.1 kOhm. 1.048 RESF * S 1.049 RNG 2kZ 1.050 RESF 1.0000kZ S 2W 1.051 DISP Adjust R27 for a reading of 1.0000+/-0.0001 kOhm. 1.052 RNG 20MZ 1.053 RESF 10.000MZ S 2W 1.054 DISP Adjust R29 for a reading of 10.000+/-0.003 MOhm. 1.055 RESF * S 1.056 HEAD AC VOLTAGE CALIBRATION ADJUSTMENT 1.057 DISP Simultaneously press the ACMA and DCV function switches. 1.058 RNG 1200V 1.059 DISP Adjust R72 for a reading of -20.0 to -30.0 1.060 RNG 2V AC 1.061 5100 1.9000V 500H S 2W 1.062 DISP Adjust R50 for a reading of 1.9000+/-0.0001V. 1.063 5100 * S 1.064 DISP Disconnect the 5100B from the UUT. 1.065 DISP Connect the 5205A and UUT as follows: 1.065 DISP [32] 5205A OUTPUT HI to UUT V-Ohm HI 1.065 DISP [32] 5205A OUTPUT LO to UUT V-Ohm LO 1.066 RNG 200V AC 1.067 5205 190.00V 50kH S 2W 1.068 DISP Adjust C2 for a reading of 190.00+/-0.05V 1.069 5205 * S 1.070 DISP Disconnect the 5205A from the UUT. 1.071 DISP [32]Connect the 5100B and UUT as follows: 1.071 DISP [32] 5100B OUTPUT HI to UUT V-Ohm HI 1.071 DISP [32] 5100B OUTPUT LO to UUT V-Ohm LO 1.072 RNG 2V AC 1.073 5100 1.9000V 50kH S 2W 1.074 DISP Adjust C19 for a reading of 1.9000+/-0.0005V 1.075 5100 * S 1.076 RNG 20V AC 1.077 5100 19.000V 50kH S 2W 1.078 DISP Adjust C10 for a reading of 19.000+/-0.010V 1.079 END